![Valco baby Vici TCD-3 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/valco-baby/vici-tcd-3/vici-tcd-3_instruction-manual_849548034.webp)
30
Appendix A: Digital I/O Control
The TCD-3 controller provides digital signals for monitoring and control of the
system. Note that system parameters such as detector temperature and filament
power levels will still require a serial link to set. After being set, the TCD-3 can act
in a stand-alone mode.
Digital IO Port
The Digital IO port is on the rear panel of the TCD-3 controller. This 8-position
port (2x4) contains three digital input signals and one digital output signal.
Pin
Signal
Pin Signal
1
Digital Input 1
5
Input Signal Ground
2
Digital Input 2
6
Input Signal Ground
3
Digital Input 3
7
Input Signal Ground
4
Digital Output – contact A
8
Digital Output – contact B
Table 9
: Digital I/O port pin assignments
Digital IO Connector
The TCD-3 system includes a removable plug for use with the Digital I/O port,
equipped with attached 4-foot wire leads. The 8-position plug uses spring clamp
connectors. If the need arises to change wire connections, use a small, flat-blade
screwdriver to insert or remove the leads.
Figure 15
: Digital I/O port location and pin numbering
Содержание Vici TCD-3
Страница 2: ...This page intentionally left blank for printing purposes ...
Страница 42: ...38 Detector Performance Log ...