4.2 Calibration of Presetter
4.2.1 Set the camera’s working distance
The working distance is defined as the distance between the front edge of the cam-
era tube shield and the object to be measured, for example a tool cutting edge.
This measurement is preset by the design and the production tolerances of the cam-
era tube shield (lens).
The working distance must be set following the instructions in the
enclosed calibration log with a permissible tolerance of 100 ±
0.1mm. (3.937 ± .004”)
Object to be
measured
Working distance
Illumination
Optical axis
Camera
Aligning the camera and the lamp