digiset
5
Urma AG, Switzerland www.urma.ch
page 16
4.2.4
Adjust X- and Z-Parallelity
You access this operating mode using the following menu steps:
Main menu
Sub-menu 1
Sub-menu 2
Setup
Ö
Corrections
Ö
Parallel correction
The axis of the calculated correction values is
automatically allocated by the system with parallel
correction.
X-axis parallelity
1. Insert calibration bar into the Digiset spindle
2. Switch on the Digiset
3. Move the measuring arm in both axis
4. Move the measuring arm to the upper position (A)
5. Click
SETUP
on the menubar;
Æ
Corrections
Æ
Parallel correction
6. The window “Parallel correction” must show “on” for
Active. Otherwise change it
7. Click on “Record”
8. the screen shows “1. Point”
9. Measure the first point (A) and click OK
10. Move the measuring arm to its lower position (B)
11. Measure the second point and click OK
12. click OK to complete the procedure.
The parallelity for the Z-axis is installed in the factory.
A
B