Software
Manual CWD 2005
___________________________________________________________________
54
Manual CWD2005.doc
6.2.1.3.1.
General
Change signal after hold provides in a soft transition to and from the calibration cycle
or when switching off the output signal hold module. This prevents a sudden rise or
fall in the measured value occurring in output signal. The transition between values is
shown in seconds and this allows the signals to blend together for a smooth transfer.
“Purge time” purges the unit with air from the fan prior to ignition. The delayed
shutdown of the air fan following a shut down removes any residual heat from the
system and the shutdown of the fan saves energy and filter life. A value of “0” will run
the cooling air fan continuously and the unit is ready for immediate use after a restart.
When selecting “Delayed shutdown” the air fan can be switched off after the time
shown in seconds. In the example, the air fan will run for a further 30 seconds after
process gas shutdown before the fan shuts down automatically. (With a delayed
shutdown value of 0 the air fan to run continuously).
“Time until power down display” determines the time before entering the screen
saver mode after a period of keypad inactivity.
“Display speed” is the constant response speed time. This value is preset at the
factory and varies from gas family to gas family and measuring ranges.
“Carrier gas Cal. cycles” determines the carrier gas calibration cycle in minutes.