Technical Data
Technical Data
System CWD 2005:
Process computer: PC104, 486, 66MHz, 8MB disk on chip flash memory.
Real time processing system with multi-tasking ability
Measuring Range: 40 to 100% F.S.
Dual Range: 20 to 100% F.S.
Low Range: 75 to 150 BTU/scf (700 to 1400 kcal/Nm3.
High Range: 1600 to 3200 BTU/scf (15000 to 30000 kcal/scf).
Accuracy: + 1.0% F.S. CV or Wobbe Index.
Specific Gravity: + 0.8% F.S. Specific Gravity.
Linearity: + 0.2%
Repeatability: + 0.5%
Response Time: 50% 7 seconds: 90% 30 seconds: 99% 45 seconds.
Outputs: 4-20 mA DC isolated (600 ohms max load).
Up to 7 mA outputs, 1 or 3 normally supplied.
RS232.
Relay output:
8 relays.
Contact rating: 1.0 amp at 100 volt.
Digital output: 4 - Open collector.
Digital Input: 3 Contact closures.
Communication: RS232.
System Details;
Gas Pressure: 6 to 15” H2O (15 to 37mbars).
Gas Consumption: 0.5 to 7 scfh (15 to 200 l/ph) application dependant.
Gas Connection: Tube fitting (Compression ¼” or 6mm).
Ambient Temperature: 5 to 40C (Rate of 5C per Hr).
Power: 115 volt 50/60Hz (330 VA) or 220 volt 50/60 Hz (165VA).
D
imensions: 30”W x 40”H x 13”D.
Weight: 111lb (50.5 kg).