Software
Manual CWD 2005
___________________________________________________________________
44
Manual CWD2005.doc
Fr = 42.01 controlled frequency of the fan
10. Stab=0.002 Stab = stability, standard deviation of the measured value
average over 2 min.This information is important for a successful calibration. A
typical value at the end of calibration is less than 0.015. Under this value
calibration is finished.
The free area of the info field is filled up with useful information in different menus for
example filter changing over temperature etc.
6.2.1. Options
The Options menu contains all of the following menus: Button
leads back to the
main menu.
The free area of the info field is reserved for important information. The information
belongs to the actual menu level. Filter change or over temperature are common
information.
Communication error is an information that the connected printer can not
communicate with the system.