Software
Manual CWD 2005
___________________________________________________________________
46
Manual CWD2005.doc
Signal is a multiple choice scrolling list field with the following text entries:
Calorific Value; Wobbe, < Wobbe >, Specific Gravity, CV net <or gross>.
Type is a list field with the following entries: 0
– 20 mA or 4 – 20 mA
”Unit” is a list field with the following contents:
Calorific: BTU /ft
3
, kJ/Nm
3
, MJ/Nm
3
, kWh/Nm
3
, kcal/Nm
3
, kJ/Sm
3
, MJ/Sm
3
, kWh/Sm
3
kcal/Sm
3
Specific Gravity: specific to air (1.0), absolute kg/Nm
3
“Range from” – “Range to” are value fields for numerical inputs.
6.2.1.1.2.
Relay output
The relay output configuration menu consists of signals in combination with threshold
values or operating status messages. If an alarm is triggered a message will appear
in the information field and the fault LED on the door will light.
Signal is a list field. Wobbe, Calorific Value and alarm messages can be selected.
Operation will show < > if an alarm value is programmed.
Unit and Value 1 and Value 2 fields contain configurable operating data.
Note:
Zero position low = no power applied to relay coil in normal operation.
Zero position high = power applied to relay coil normal operation.
Eight normally open or normally closed contacts are available. (See section 10.2.1).