1.1.
Safety Instructions
Before start up and use of the CWD 2005, please take time to carefully review this
manual and its contents. Observe all Warnings, Cautions and Notes.
Warning:
Reasons to safe persons allow maintenance only after disconnect the main switch.
Before maintenance or repair work on the electrical or on the high voltage equipment
the following points have to be considered.
After switching off the main switch the capacities of the equipment store up to 5 min
high voltage. It is important to wait this time till the voltage brake down to zero.
At the installation of the protective earth please consider the existing rules for
instruments with higher leakage current (> 1,5 mA)
Please pay attention about the safety function of the person current circuit breaker. It
is forbidden to manipulate to get a malfunction. (safty switch in the main door).
Before the service at the burner system consider that parts of the burner could be
very hot (more than 200 °C). Please wait more than 15 min to cool down all burner
parts before doing any service.
At the start up and service or not optimal conditions (not 100 % burnt gas ) there is a
danger of bad components in the burnt gases of the outlet of the system.
If process gas has toxic components please consider that all actual safety rules have
to be abode.
Non toxic gases are without problems. They are very high diluted with cooling air.
All electrical connections to line power must be carried out by a qualified electrician.
The analyzers internal electrical components are wired in accordance with NEC, VDE
and International standards. Check the power supply voltage, details can be found on
the technical data label inside the door.
The gas connections have to be installed only from fort his work qualified persons. All
for this job actual laws and rules have to be considered.
The CWD 2005 operates over a specified ambient temperature range. Installing the
instrument directly outside is not recommended, since damage to the analyzer may
occur from the adverse weather conditions if an outdoor shelter is not used.
Caution:
Caution is necessary during installation because pressurized flammable gases are
connected to the instrument and any installation work performed by the customer
must conform to all applicable local and national codes. All internal sample gas
components have been leak tested at 1.5 times the normal operating pressure during
assembly at the factory. After installation, the entire sample gas system should be
pressurized and tested for leaks before use.