4. Operation 

4.1 Caution on Operation 

 

 

WARNING 

 

This pump is not pressure-proof. 

Never  run  the  Vacuum  pump  on  blocking  up  the  exhaust  outlet, 

putting any device that hampers the gas passage or putting any oil 

mist trap not conforming to the ULVAC pump. There is a risk that the 

pressure inside the rises up as far as the Pump unit and/or Oil level 

gauge breaks or the Motor gets overloaded. 

 

This  product  is  not  made  as  the  withstand  pressure  structure. 

Ensured pressure value of the Pump shall be 0.03MPaG (0.3kg/cm2

G

(Gauge pressure). 

 

If  any  valve  was  put  to  a  pipe  after  the  Exhaust  outlet,  check  and 

ensure that it is open. 

 

 

CAUTION 

 

1) The Pump oil might deteriorate in a shorter time depending on the 

use. 

It  is  recommended  to  replace  the  first  Pump  oil  within  ten  days 

after operation start and see how it got dirty to determine the oil 

replacement cycle. 

 

2)  If  the  Pump  breathes  in  a  lot  of  water  or  the  like,  you  should 

replace  the  oil  more  frequently.  If  kept  operation  without  getting 

rid of breathed water, it would deteriorate lubrication of the oil and 

further help corrosion of the Pump inside and result in causing a 

failure. 

Do not store the product keeping sucked the water. 

 

3)  If  the  Pump  breathed  in  chemical  material  such  as  acid, 

immediately replace the oil as it would cause the rust during the 

stop in one night to make the system e not applicable to operate. 

We shall be not liable to the durability to chemical material. 

 
 

Содержание VD151

Страница 1: ...2 08 INSTRUCTION MANUAL OIL SEALED ROTARY VACUUM PUMP Before using this product be sure to read this operation manual Keep this manual with care to use at any time ULVAC Inc Components Division http w...

Страница 2: ...e ensured if any of the devices put together did not conform to same regulations or the product itself was modified ULVAC shall be not liable to guarantee performance and safety in such cases above An...

Страница 3: ...makes a mistake in handling it indicates a possibility that the user is subject to death or heavy injury CAUTION If the user makes a mistake in handling it indicates a possibility that the user is su...

Страница 4: ...active gas is not allowed as there is a risk of leakage of the gas from the Pump unit if it was exhausted by the vacuum pump DANGER Use of the toxic combustible or combustion susceptible gas and subst...

Страница 5: ...r at the top right of the Instruction Manual cover and issue the revised version WARNING This Instruction Manual shall absolutely need to be delivered to the last user that uses the product WARNING To...

Страница 6: ...erating the unloading machinery 3 There is a risk that the Pump might drop or lay down when attempted unreasonable operation or machinery setup was not sufficient You are strictly restricted to enter...

Страница 7: ...sure that any of hazardous energy is blocked before starting the operation 2 Entitled staff should conduct the wiring operation Erroneous wiring work might cause a fire 3 Conduct the wiring operation...

Страница 8: ...us atmosphere by explosive gas It might cause injury and or fire WARNING 1 Be sure to turn OFF the Power Supply to execute check and repair You have a risk of getting electrical shock or injury by acc...

Страница 9: ...ou have a risk of getting burned or fire caused by over heat Power Supply wiring Dismantling WARNING Be sure to cut off the electricity before starting install or dismantling operation Inlet piping Di...

Страница 10: ...ture Ensured pressure value of the Pump shall be 0 03MPaG 0 3kg cm2G Gauge pressure Do not run the Pump on blocking the Exhaust outlet or putting any device that might hamper gas passage onto the outl...

Страница 11: ...YK08 0028 DI 002 08 x 6 INLET 7 OUTLET...

Страница 12: ...YK08 0028 DI 002 08 xi Fig 1 Warning Label VD151 VD201...

Страница 13: ...ter beneath the Pump 4 Since you may get injury due to an anchor being used in the box guide the worker to wear the leather gloves and to use the appropriate tool Upon delivery of the product check fi...

Страница 14: ...on 2 Ambient temperature and humidity for operation 4 C to 40 C less than 80 RH no condensation However ULVOIL R 42 shall be used for 4 10 C operation and ULVOIL R 72 shall be used for 10 40 C operati...

Страница 15: ...pped with the protective device Put an overload protective device to connect through the motor with the Power Supply It is imperative to put the overload protective device under direction of the Elect...

Страница 16: ...and Safety Measures 1 1 1 1 Danger Leakage of dangerous gas and dangerous materials 1 1 1 2 Warning Transfer of heavy material 1 1 1 3 Warning Electric shock 2 1 1 4 Warning Explosion 2 1 1 5 Caution...

Страница 17: ...oise and Abnormal Vibration 33 6 2 7 Checking The Coupling and Spider 33 6 2 8 Checking The Oil Mist Trap 34 6 2 9 Checking the drain valve 35 6 3 Checkup after storage for a long period 35 6 4 Overha...

Страница 18: ...Main displacement parts list VD201 46 Fig 1 Warning Label xi Fig 2 VD151 Dimensional drawing 6 Fig 3 VD201 Dimensional drawing 7 Fig 4 Lubrication and oil level indication 9 Fig 5 Connection to vacuu...

Страница 19: ...ial or sucked substance at the occasion of check or disposal 1 When you check the pump please wear a brace that supports the toxic substances that exhaust the pump 2 To overhaul or dispose ask the spe...

Страница 20: ...d pressure value of the Pump is 0 03MPaG 0 3kg cm2 G Gauge pressure Check the Exhaust side pressure of the pump If it was over 0 03 MPaG 0 3 kg cm2 G Gauge pressure take away anything in and around th...

Страница 21: ...in need Read it with attention to acknowledge the toxic characteristics described on the MSDS Please do not use the chemical material except the above mentioned chemical material vacuum pump oil CAUT...

Страница 22: ...stopped for a long time without venting the intake side to atmospheric pressure the pump oil will flow back to the cylinder making restart difficult This pump is equipped with the oil anti sucking fe...

Страница 23: ...is the value indicated on the Pirani gauge when the standard oil ULVOIL R 72 is used The ultimate pressure when the McLeod gauge is used is approx one digit lower than the value indicated on the Pira...

Страница 24: ...YK08 0028 DI 002 08 6 Fig 2 VD151 Dimensional drawing...

Страница 25: ...YK08 0028 DI 002 08 7 Fig 3 VD201 Dimensional drawing...

Страница 26: ...he machine horizontal to a place where there are less dust and humidity Make a layout taking into consideration of works such as setting removal check cleaning and so on As for the environmental condi...

Страница 27: ...ation Should the oil touched to your hand are entered in your eye immediately follow the emergency treatment described on the MSDS CAUTION Ensure to use the vacuum pump oil designated by ULVAC Operati...

Страница 28: ...pump is not transmitted to the vacuum chamber 3 Provide a vacuum valve vacuum gauge and vent valve between the vacuum chamber and pump as shown in Fig 5 Fig 5 Connection to vacuum chamber example WAR...

Страница 29: ...p to cause break or oil leak of the casing or Oil level gauge resulting in overload of the motor WARNING 1 Be sure to use products made of electricity conducting material for the Duct piping Should yo...

Страница 30: ...fety circuit such as Electromagnetic breaker for the electrical connection WARNING Install and operate the product in compliance with the laws and regulations relating to the safety e g Fire Defense L...

Страница 31: ...0V 50Hz 1 90 400V 50Hz 2 00 415V 50Hz 2 00 415V 50Hz 1 50 380V 60Hz 1 50 380V 60Hz 1 60 400V 60Hz 1 60 400V 60Hz 1 60 440V 60Hz 1 60 440V 60Hz 1 70 460V 60Hz 1 70 460V 60Hz 3 60 200V 50Hz 3 60 200V 50...

Страница 32: ...YK08 0028 DI 002 08 14 Fig 6 Terminal box internal wiring diagram Fig 7 Electrical wiring diagram...

Страница 33: ...the terminal box with 380V 460V and 200V 240V power supply voltage please change the wiring WARNING 1 Be sure to put an overload protective device conforming to the motor capacity You have a risk that...

Страница 34: ...sensor PTO N R T Nominal running temperature of the PTO WARNING 1 Be sure to use different power cords for the motor and for taking out a temperature sensor signal 2 Apply a voltage of 250 V or less...

Страница 35: ...that it is open CAUTION 1 The Pump oil might deteriorate in a shorter time depending on the use It is recommended to replace the first Pump oil within ten days after operation start and see how it got...

Страница 36: ...hour or more under high pressure 1000Pa or more would increase the oil volume that are discharged as the oil mist and cause rapid parts wear or cause a trouble such as burning You are recommended to...

Страница 37: ...After checking 1 2 and 3 above close the vent valve and run the pump Here ensure that the vacuum gauge between the vacuum valve and the pump indicates a pressure close to the ultimate pressure WARNING...

Страница 38: ...ight enter inside the Pump cylinder if you left the Pump stopped longtime three days or more even you kept the atmospheric pressure inside the Pump when stopped it last time The Overload protective de...

Страница 39: ...r non load operation 20 temperature increase under high load operation 78 Refrain from touching the Motor and Pump unit until the Pump cools down after having stopped operation CAUTION 1 Ensure to clo...

Страница 40: ...ow temperature shall be lower than the specified process performance Note further that keeping the Gas ballast valve open when not breathing in the condensed gas shall cause the Pump oil splashing and...

Страница 41: ...is the overload phenomenon caused by increased viscosity of the pump oil To make sure you should confirm that the capacity of the motor overload protective device conforms to the motor rated value the...

Страница 42: ...pump instead of the original exhaust pipe at the outlet 1 At the outlet of the pump remove the original exhaust pipe 2 You assemble the adapter and TMX 1 to establish the pump 3 Between the adapter a...

Страница 43: ...raised oil mist trap please exchange We recommend you to install a Pressure monitor To use the Oil mist trap be cautious not to have clog of the filter in the trap Too much clog would prevent the exh...

Страница 44: ...n the site shall likely to raise the ultimate value higher than the catalogue value as the vacuum gauge was put far away from the Pump or maybe affected by the water drop rust or other substances atta...

Страница 45: ...mping speed MODEL VD151 VD201 0 5 10 15 20 25 0 1 1 10 100 1000 10000 100000 Pumping Speed m 3 hour Pressure Pa Gas ballast closed Gas ballast opened Pumping Speed JIS B 8316 1985 Type VD151 VD201 Pow...

Страница 46: ...hile the suction pressure was between 3 104 4 104 Pa Operation opening the Gas ballast valve would require larger motive energy at all the time as its compression work is large even the suction pressu...

Страница 47: ...nts depending on the use of the Pump you should check the following regularly it is helpful to avoid trouble and extend the pump life cycle WARNING 1 Be sure to turn OFF the Power Supply to execute ch...

Страница 48: ...iction parts of the pump In the worst case the pump will seize up and cannot be rotated Table 4 gives a guide for pump oil change frequency CAUTION VD201 The oil level gauge is for checking the pump o...

Страница 49: ...4 You should also replace the oil if breathed in the solution to deteriorate lubrication of the oil as it would also cause biting inside You shall have a risk if breathed in the solution in operation...

Страница 50: ...armitupandthenchecktheultimatepressure 5 If the specified ultimate pressure cannot be attained by oil change sludge or other deposit may have collected on the bottom of the pump case In that event ove...

Страница 51: ...p 1 Check whether bolts and nuts and the like fixing the pump are loose or not 2 Check whether pipes connected to the inlet outlet are loose or not 3 Check and ensure that there is no leakage from the...

Страница 52: ...ing out the Motor Fig 11 Replacement of coupling spider 6 2 8 Checking The Oil Mist Trap To use the Oil mist trap be cautious not to have clog of the filter in the trap Too much clog would prevent the...

Страница 53: ...gardless of the check items described above Overhaul shall be required to keep the performance as well as the safety and further to continue the production on forecast Please contact the Service Cente...

Страница 54: ...d the reactive gas 6 The pump is not connected to the power supply 7 The power switch is not turned on 8 Problem with power supply voltage 9 The overload protector has actuated 1 Check the connection...

Страница 55: ...is low 15 At the pump stop Not to vent the pump to atmospheric pressure 10 Replace the motor 11 Overhaul replace the cylinder and rotor 12 Overhaul replace the vanes 13 Overhaul replace the damaged c...

Страница 56: ...ign matter has entered the pump 9 Components inside the pump have burnt out 1 Do the connection again to correct the rotation direction 2 Check the Cooling fan mount and clear the error 3 There is no...

Страница 57: ...is clogged b Oil distributor valve has a trouble 11 Water entered inside the pump 1 Select another Pump type 2 Measure correctly the pressure 3 Use the Vacuum gauge that matches the measurement pressu...

Страница 58: ...e would rise up around 100 on continuous operation under high suction pressure No problem 2 Control the oil level Supply the oil by the specified volume 3 Install a cooling device such as the Gas cool...

Страница 59: ...d Oil seal Replace them if necessary 2 Tighten again the Oil inlet 6 2 3 6 4 Abnormal Motor current value 1 Foreign substance entered inside the Pump impaired the Motor rotation 2 Abnormal sliding of...

Страница 60: ...ny toxic gas Note that you are requested to bear the cost and charges relating to the disposal WARNING 1 You should ask a special disposal agency for the disposal particularly if the Pump has exhauste...

Страница 61: ...standard specifications although the product is used under the service conditions described in this document such as temperature range and power etc 2 Direct export transaction Product which has damag...

Страница 62: ...cy or secondary damnification occurred to Buyer from law suit to ULVAC by third party for patent infringement 8 The reason of the failure deeemed below the specified usage condition by ULVAC technical...

Страница 63: ...Cylinder 16 O ring S 5 NBR 1 17 O ring S 16 NBR 1 18 O ring S 70 NBR 2 19 Outlet valve 13 9 5 9 FPM 3 20 Outlet valve spring 10 20 SUS 3 2nd 21 Oil seal SC17 30 7 NBR 1 intermediate cover 22 O ring S...

Страница 64: ...spring 2 6 31 SWP A 5 1st Cylinder 16 O ring S 5 NBR 1 17 O ring S 16 NBR 1 18 O ring S 85 NBR 1 19 Outlet valve 13 9 5 9 FPM 3 20 1st Outlet valve spring 10 20 SUS 3 2nd intermediate 21 Oil seal SC...

Страница 65: ......

Страница 66: ...ty period Table Making format for names and contents of hazardous substances or elements Name of parts Hazardous substances or elements Pb Hg Cd Cr6 PBB PBDE Body indicating that content of the hazard...

Страница 67: ...ion Remarks Contaminant Check an applicable box I guarantee that above returned item s is not contaminated with harmful substances Above returned item s is contaminated with the following harmful subs...

Страница 68: ...o jp en Service Centers http www ulvac co jp en support service center Sales O ces http www ulvac co jp en support sales o ces 253 8543 TEL 0467 89 2261 108 0075 TEL 03 5769 5511 532 0003 TEL 06 6397...

Отзывы: