0.5.2 Transfer
WARNING
Pump Weight
VD151
:
29kg
VD201
:
34.5kg
When you carry a pump, you have a risk of giving damage to your back.
Be sure to use the loading machinery (such as mobile crane) to lift up
the Pump or load it on the pallet and fix it and run the Pallet truck for its
transfer.
0.5.3 Ambient Condition for Storage, Install and Operation
As precise clearances are provided with this machine, be sure to fulfill the following for its
storage, install and operation;
(1) Ambient temperature and humidity for storage
: -10°C to 60°C, less than 95%RH
(no freeze, no condensation)
(2) Ambient temperature and humidity for operation : 4°C to 40°C, less than 80%RH
(no condensation).However, “ULVOIL
R-42” shall be used for 4-10 °C operation
and “ULVOIL R-72” shall be used for
10-40 °C
operation.
(3) Height (for both storage and operation) : Lower than el. 1,000m.
(4) External vibration (for both storage and operation) : Vibration acceleration less than
114dB (0.5G)
(5) Miscellaneous (for both storage and operation)
:
a. There shall be no corrosion behavior or explosive gas.
b. There shall be no freeze or dew formation.
c. There shall be no dust.
d. It shall be in house.
e. Another pump shall not be put on the Pump.
The Pump shall not be laid down nor put touching its motor edge face or oil gauge edge
face with the ground.
f. There shall be no direct sun beam.
g. Heat source shall be put away from the Pump.
Содержание VD151
Страница 11: ...YK08 0028 DI 002 08 x 6 INLET 7 OUTLET...
Страница 12: ...YK08 0028 DI 002 08 xi Fig 1 Warning Label VD151 VD201...
Страница 24: ...YK08 0028 DI 002 08 6 Fig 2 VD151 Dimensional drawing...
Страница 25: ...YK08 0028 DI 002 08 7 Fig 3 VD201 Dimensional drawing...
Страница 32: ...YK08 0028 DI 002 08 14 Fig 6 Terminal box internal wiring diagram Fig 7 Electrical wiring diagram...
Страница 65: ......