6. Maintenance
The chapter describes necessary maintenance so the product can be used for years to come. If you
neglect maintenance, the product will wear down more quickly which may cause the product to stop
working. Please read this chapter carefully and perform maintenance.
6-1. RF power supply maintenance and inspection
RF power supply maintenance and inspection is described below. About once every six months,
remove the RF power supply's cooling water tubes and inspect the connectors for corrosion and
cloging. Never open the cover when energized.
The parts required for the RF power supply's periodic maintenance are listed below.
Table 6-1 RF power supply recommended replacement parts
No.
Part name
Periodic maintenance -
Recommended replacement
period
Comments
1
Cooling fan
4 to 5 years
2
Electrolytic
capacitor
4 to 5 years
3
DC/DC converter
4 to 5 years
4
Switching power
supply
4 to 5 years
Power supply for internal device control
5
Fuse
4 to 5 years
6
Flow meter
4 to 5 years
7
Cooling plate
5 to 7 years
(Basically, handled by internal
washing)
The cooling plate may clog because
of the water quality.
Note: The recommended replacement period is not a guarantee of a product's lifespan. The rate of
consumption of these products may also greatly differ according to the frequency of device usage.
Note: In order to use this product for many years to come, please ask ULVAC to periodically overhaul
the product using the above table as a reference.
Warning
When performing maintenance, always shut off the primary-side breaker of the power
supply used and pull the plug from the input socket before working. Also wait for three
minutes or more after shutting off the breaker.
Danger
Never open the power supply's cover when energized.
The voltage generated by this
power supply is sufficient enough to pose a danger to human life. If by some chance a
person touches this current, it will lead to a serious accident.
Never energize the power
supply with the cover removed.
Содержание RFS-13100F
Страница 66: ...F Power Supply and Matching Box 58 Figure 8 1 RFS 13100F...
Страница 67: ...F Power Supply and Matching Box 59 Figure 8 2 RFS 13150F ESC ENT M8...
Страница 68: ...F Power Supply and Matching Box 60 M8 60Kg Figure 8 3 RFS 13200F...
Страница 69: ...F Power Supply and Matching Box 61 Figure 8 4 RFS 27400F...
Страница 70: ...F Power Supply and Matching Box 62 Figure 8 5 MBX 13100F...
Страница 73: ...F Power Supply and Matching Box 65 Figure 8 8 MBX 27400F 2 M6...