30
YO20-0938-DI-001-00
●
In addition, a high-frequency leakage current occurs as an inverter is used inside. When using a
ground-fault interrupter other than the specified ones with measures already taken against high
frequency, use one with a sensed current of 200 mA or more.
Table 8:
List of power capacities and ratings of recommended ground-fault interrupters
Rating List
200V
400V
Recommended type
NV-32SV(F)-
(
Manufactured by Mitsubishi Electric
)
Rated current
(
A
)
30
15
Rated sensed current
(m
A
)
30
Power capacity
(
kVA
)
9.5
■
Pin assignment of power
Figure 1:
Pin assignment
Pin assignment
Pin
No.
Specifications
A
Phase L1/R
B
Phase L2/S
C
Phase L3/T
D
PE/GND
Table 9:
Receptacle connector specifications
Manufacturer of receptacle
connector
DDK Ltd.
Japan Aviation Electronics Industry, Ltd.
Receptacle connector type
CE05-2A22-22PD-D
JL04HV-2E22-22PE-B-R
Applicable plug type
(Direct type)
CE05-6A22-22SD-D-BSS
JL04V-6A22-22SE-EB-R
Applicable electric wire size
Up to 10AWG/4
(
Outer diameter of insulation coating:
φ
4.5 mm
)
Rated current (50/60 Hz)
46A
After connecting the power connector, install the guard.
Figure 14:
Installation diagram of power connector guard
Power connector
Power connector guard
Screw
Содержание MS120A
Страница 2: ......
Страница 10: ......
Страница 14: ......
Страница 47: ...33 YO20 0938 DI 001 00 Figure 16 I O block diagram for remote control wiring...
Страница 65: ...51 YO20 0938 DI 001 00 7 2 External Dimensions Figure 17 External dimensions of MS120A MS120A...
Страница 67: ...53 YO20 0938 DI 001 00...
Страница 68: ...54 YO20 0938 DI 001 00...
Страница 69: ...55 YO20 0938 DI 001 00...