CRYO-U
®
Cryopump Instruction Manual
ULVAC CRYOGENICS INCORPORATED
A-3
Problem
Possible cause
Corrective action
Ⅲ.
(Continued)
A
compressor
unit
starts, but a cryopump
doesn't start. Or the
cryopump is stopped,
but the compressor
unit
is
running
continuously.
3) The drive motor of a refrigerator
unit has mechanical damage.
①
Motor shaft failure or bearing
defects.
②
Drive motor circuit failure.
Check whether there is sound which
shows a motion of a motor by turning on
and off the switch of the refrigerator.
When no sound or irregular beat comes
from the motor, contact our Service
Engineering Division or the nearest CS
center.
Measure resistance and insulation
resistance between the drive motor
connector pins (Impressed voltage: DC
500V) and inform us the measured
values with S/N of the refrigerator unit.
Ⅳ.
The cryopump fails to
cool down to 20K or
lower at the time of
startup.
1) Helium Supply/Return line
incorrectly connected.
Self-sealing coupling is not
fully tightened.
Properly connect the flexible hoses
referring to Sec. 4.7.
Check that all self-sealing couplings
are fully connected referring to Figure
4-5.
2) Pressure in the cryopump is too
high.
Make sure that there is no leakage.
The cryopump may fail to cool down
due to thermal load of the gas caused
by a leakage
.
3) Heat load to the cryopump is
too high.
If the temperature of the cryopump
rises when opening the main valve,
heat input amount from the user
’s
device is over the acceptable value. In
this case, take measures to reduce the
heat load.
If the cryopump still fails to cool down
to 20K or lower even when the main
valve is closed, the refrigerator unit
may be out of order. Contact our
Service Engineering Division or the
nearest CS center.
Содержание CRYO-U12H
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Страница 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 46: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 60: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Страница 64: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 78: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 86: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 88: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 100: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...