CRYO-U
®
Cryopump Instruction Manual
ULVAC CRYOGENICS INCORPORATED
8-9
Figure 8-2 Disconnecting Flexible Hoses
5.
Reduce the pressure in the refrigerator unit and the flexible hoses to a level of 0.3MPaG by
opening the valve(V1) of the charging adaptor.
NOTE:
As refrigeration lowers the temperature of helium gas in the refrigerator unit, the density of
the gas increases. After shutdown of the refrigerator, the temperature of the helium gas in
the refrigerator returns to room temperature from cryostatic temperature gradually, and the
inner pressure rises at the same time. The unit has a pressure relief valve which opens at
the inner pressure of 1.9MpaG and more. If once the pressure relief valve opens, it could
cause a leakage that would depend on the surrounding environment such as dust.
Therefore, ensure to reduce the inner pressure right after shutdown of the refrigerator in
order not to work the pressure relief valve.
Figure 8-3 Connecting Charging Adopter
OUT
IN
SUPPLY GAS
COLD HEAD POWER
NO.1
NO.2
NO.3
INPUT POWER
RESPONSE
REMOTE
RETURN GAS
RETURN GAS
WATER IN
WATER OUT
RETURN
SUPPLY
Cryopump
Coldhead Cable
Disconnect here
Power Cable
Cooling Water Line
Flexible Hose
SUPPLY
RETURN
V3
V1
V2
D-P
RO
D-P
RO
-
0.1
3.5
3
2.
5
2
1.5
1
0.
5
0
Cryopump
Charging Adopter
Flexible Hose
Coldhead Cable
Release He gas
Release He gas
Содержание CRYO-U12H
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Страница 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 46: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 60: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Страница 64: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 78: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 86: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 88: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 100: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...