CRYO-U
®
Cryopump Instruction Manual
ULVAC CRYOGENICS INCORPORATED
8-13
12.
Confirm that the pressure gauge on the front panel of compressor unit indicates the
specified pressure (indicated on the front panel). If not, charge helium gas through the
charging valve of the compressor unit again, or adjust the pressure by discharging the
helium gas.
13.
Restart the cryopump (refrigerator unit).
Figure 8-6 Connecting Flexible Hoses
OUT
IN
SUPPLY GAS
COLD HEAD POWER
NO.1
NO.2
NO.3
INPUT POWER
RESPONSE
REMOTE
RETURN GAS
RETURN GAS
WATER IN
WATER OUT
RETURN
SUPPLY
Cryopump
Coldhead Cable
Connect Here
Power Cable
Flexible Hose
Cooling Water Line
Содержание CRYO-U12H
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 24: ...This page intentionally left blank S 12 ULVAC CRYOGENICS INCORPORATED...
Страница 26: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 46: ...Installation 4 10 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 60: ...Regeneration 6 8 ULVAC CRYOGENICS INCORPORATED This page intentioanlly left blank...
Страница 64: ...Disconnection and Storage 7 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 78: ...Maintenance 8 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 86: ...Troubleshooting A 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 88: ...Flexible Hose B 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 100: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...