Chapter 8: DICOM
00.053.204, Revision A
SonixTouch Q+ Service Manual
8-22
To Configure AE Configuration-Specific Advanced Print Settings:
1.
Tap the touch screen
Menu
button.
2.
Select
Administrator > DICOM > Print… > AE Configuration > Settings… > Advanced Print Settings
.
3.
Complete
Advanced Print Settings
as required.
4.
Select
OK
to accept the changes or
Cancel
to exit without saving.
To Delete DICOM Print Devices:
1.
Tap the touch screen
Menu
button.
2.
Select
Administrator > DICOM > Storage…
.
3.
Under
Devices
, highlight the relevant
Name
.
4.
Select
Delete
to remove the device from the list.
5.
Repeat
and
as many times as necessary.
6.
Select
OK
to accept the changes or
Cancel
to exit without saving.
Note:
for details on these options.