SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 3: System Installation
3-11
3.3
INSTALLATION
3.3.1
Mechanical Inspection
3.3.1.1 Wheels
The system is equipped with dual lock castors that can be set in three distinct positions.
Figure 3-2: Dual Lock Brakes
Table 3-1: Dual Lock Brakes
Ensure brakes function as expected in all three positions:
• Setting
1
: system cannot move at all
• Setting
2
: system moves only back and forth (ensure it cannot move sideways)
• Setting
2
: system moves freely in any direction.
3.3.1.2 LCD Display
Tip/rotate the LCD display to ensure that it:
• tips forward and back as required
• rotates from side to side
• stays firmly in position once the required angle and rotation is achieved.
1
Total lock: casters cannot move (both levers in lowest position).
2
Directional lock: casters are directionally locked, restricting them to front/back movement only (dark gray lever in
lowest position, light gray lever in highest position).
3
Completely unlocked: casters can move freely in any direction (both levers in highest position).
Note:
The system will be more difficult to move around on carpeting due to increased friction on the casters.
1
3
2
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......