Chapter 7: Field Service Components
00.053.204, Revision A
SonixTouch Q+ Service Manual
7-4
To Import User Data from a Backup:
1.
Ensure the system is plugged in and powered on.
2.
Plug the previously-created backup into one of the four USB ports.
3.
Tap the touch screen
Menu
button.
4.
Select
Administrator > System > Import…
.
5.
Check the items to be imported.
6.
Select
OK
to begin the import process or
Cancel
to exit without importing.
Caution:
Ultrasonix does not recommend importing user-defined
Presets
created with a previous software version
as they may not be compatible for use with a more recent software update.
Note:
Use
Select All
to check all items and
Clear All
to clear all checkboxes in one step.
Only active
Protocols
with changes to default
Presets
,
Measurement
and
Worksheet
settings will be
available for Import.