Chapter 4: Performance Testing
00.053.204, Revision A
SonixTouch Q+ Service Manual
4-2
4.2
IMAGE TEST
To Conduct an Image Test:
1.
Connect each of the different transducers available and tap the touch screen B button.
2.
Verify that the image scanned has all elements present. Note any black static line(s) that are visible on the
image.
3.
If found, test with the other transducers to verify the non-visible element(s) is in the same location.
4.
Try different transducer connectors to eliminate the possibility of a malfunctioning transducer.
4.3
REMOVABLE (PERIPHERAL) HARDWARE (IF APPLICABLE)
The system supports many different types of removable hardware other than what comes with the system as a default
package. The following comprises a list of the supported removable hardware:
• medical grade USB printer
• medical grade footswitch (dual or triple)
• barcode reader
• USB media (memory stick, external hard drive, etc.).
• extra (HDMI-compatible) LCD display (
4.4
UPS TEST (IF APPLICABLE)
To Test the UPS:
1.
Ensure that the UPS has been configured (
).
2.
Ensure the system is plugged in and powered on.
3.
Unplug the system's power cord from the wall.
4.
Ensure the system is still running (i.e., the system is running on battery power).
5.
Plug the system back in.
6.
Power off the system.
Note:
Most transducers can be inserted in any of the transducer connection ports. However, to ensure proper
function insert:
• the SA4-2/24 transducer into the second or third transducer connection port
• 4D transducers into the first (upper-most) transducer connection port.
Only one 4D transducer can be connected at a time.
Caution:
DO NOT plug the extra (HDMI-compatible) LCD display into the peripheral receptacle (
). Refer to
for instructions on connecting/installing the removable (peripheral) hardware.
Note:
To ensure the battery is completely charged at the time of installation, leave the system plugged in for at
least 6.5 uninterrupted hours.
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......