Chapter 7: Field Service Components
00.053.204, Revision A
SonixTouch Q+ Service Manual
7-18
6.
Disconnect the three (bottom left rear) power cables connected to the modulo: LCD display power cable (
1
),
UPS input power cable (
2
) and UPS output power cable (
3
).
7.
Using the 4mm Allen key, undo the two screws and washers securing the modulo to the front of the system
frame.
8.
Using the #2 Phillips screwdriver, undo the four screws securing the modulo to the side of the system frame.
9.
Pull the modulo out of the system frame, then lift it clear.
Note:
The LCD display power cable (
1
) is color coded with
green
tape.
System Power Connection