Chapter 8: DICOM
00.053.204, Revision A
SonixTouch Q+ Service Manual
8-12
8.1.4
DICOM Worklist Settings
DICOM Worklist Settings
offer advanced settings for configuring the
DICOM Worklist Service Class User (SCU)
.
Figure 8-12: DICOM Worklist Device Test Results
To Access DICOM Worklist Settings:
1.
Tap the touch screen
Menu
button.
2.
Select
Administrator > DICOM > Worklist
.
Note:
All
Modality Performed Procedure Steps
(
MPPS
) functionality is invisible to the user, except when testing
a
DICOM Worklist Device
.
MPPS
will be accepted or rejected based on the
DICOM Server’s
settings.