Chapter 3: System Installation
3-19
3.3.5
Transducer Inspection
To Inspect the Transducers Prior to Connection:
1.
Unpack each transducer and inspect for damage (e.g., cracks in the case or bent pins).
2.
Note the model and serial numbers of any damaged transducers and report them to Ultrasonix.
3.
Ensure that all transducers can be connected to each connection port.
3.3.6
System Initialization
To Initialize the System:
1.
Inspect the system for scratches or damage. Note any damage to the system and report it to Ultrasonix.
2.
Connect at least one transducer to the system.
3.
Plug in the power cord.
4.
Press and hold the console
POWER
button for one second.
5.
Inform the facility's representative that the system is installed and ready for any safety testing they would
normally conduct.
6.
Once the system is running, the LCD display will show a black background image field with settings
information or the
Exam Management
page.
Note:
Most transducers can be inserted in any of the transducer cone ne ct ion ports. However, to ensure
proper function insert:
• the SA4-2/24 transducer into the second or third transducer connection port
• 4D transducers into the upper-most transducer connection port.
Only one 4D transducer can be connected at a time.
Note:
If equipped with a UPS, during installation the UPS battery will need to be completely charged. Plug in
the system and let it charge uninterrupted for at least 6.5 hours.
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......