![UI SP-556 series Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/ui/sp-556-series/sp-556-series_operation-manual_3276528034.webp)
- 31 -
Mode No.
Analog output:Setting of measurement choice and the output digit (Option AV3-5/AI)
C
A B C D E
C. 1 0
Digit selection
0:Right 4digits:comparison
1:Left 4digits:comparison
Analog output method
0:Real time output
1:Synchronizes for display
2:Synchronizes for measurement (calculation value)
〔Digit selection〕
The four digits for comparison output are selected.
Right 4 digits
Left 4 digits
〔Analog output method〕
0:Real time output
The real time is analog output in sync with inside calculation.
The real-time output functions only when I set Mode No.0
“
00
”
(A input) or
“
01
”
(B input) or
“
08
”
(Passing time measurement).
Otherwise, set it in 1(Synchronizes for the display).
1:Synchronizes for the display
The analog output is output for an indication value.
When hold input functions, the analog output is output for a displayed value
now. (It synchronizes at indication sampling time)
For example when a peak hold is functioning, analog output by the present
shown value (Peak hold value).
2:Synchronizes for the measurement (calculation value)
The analog output is output for a calculated value.
(It synchronizes at indication sampling time)
The difference with 1(Synchronizes for the display), When it is input hold,
not an indication value, it is made analog output for a calculated value.