Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
31
6.3.
TROUBLE-SHOOTING
When the furnace detect a failure, a message display in the screen, and an alarm sound. Failures’ guide:
Message
Explanation
Cause
Process to follow
SENSOR
FAILURE
Thermocouple
disconnected
The temperature sensor is
defective or not properly
connected
-
stop the alarm sound: press
STOP
-
turn off furnace (switch ON/OFF)
-
contact the after-sale service
HEATING
FAILURE
Maximum
permissible
temperature is
reached
When the final temperature is
exceeded by:
·
50°C for a parameter setting
>800°C
or
·
70°C for a parameter setting
<800°C
-
the chamber opens to prevent
damage to the unit
-
stop the alarm sound: press
STOP
-
turn off furnace (switch ON/OFF)
-
wait the cooling and try again
-
if the failure persist, contact the
after-sale service
MOTOR
FAILURE
(1/2)
Tray blocked
The tray’s movement is
interrupted
-
stop the alarm sound: press
STOP
-
check that any objet don’t block
the tray’s movement, and
remove-it if necessary
-
press the RETURN key to clear
the failure on the screen
-
if the failure persist, contact the
after-sale service
MOTOR
FAILURE
(2/2)
Non-
conformed
open/close
time of the
furnace
This message is displayed if don’t
detect a movement of the tray
after 30 seconds
Probable cause: geared motor
failed or not connected
-
stop the alarm sound: press
STOP
-
press the RETURN key to clear
the failure on the screen
-
turn off furnace (switch ON/OFF)
-
contact the after-sale service