Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
22
External cooling time (0 to 60 minutes)
-
Set at 0: the chamber opens in one phase, after vacuum flooding and holding time
-
If the value is > 0: the tray is moved in 6 equal stages during the indicated time. The chamber can be
opened in one phase by press the key
Stand-by temperature (minimum 70°C)
Always ≥ night temperature (70 °C) and ≤ preheating temperature
Stand-by delay (0 to 60 minutes)
When the furnace has reached the standby temperature, it indicates the remaining time before the
controlled closure of the firing chamber. Any action on the screen will delay this operation.
Return night program
This function (selected by default) allows to raise the temperature in the firing chamber to 70°C after 3
hours in stand-by mode. When the user is absent, it limits the electricity consumption while keeping the
furnace operational.
IMPORTANT: In night mode, the stand-by screen disappears (black screen). Only the blue LED lights up.
It blinks when a cycle is terminated before entering the standby mode. If the furnace remains in
standby mode, it lights up continuously
Calibration
Each programme can be calibrated individually. This value is preset at 0 and should ensure the proper
results. However, the user can adjust this parameter based on his own assessment criteria.
This can be done in stages of one degree (between +50°C and -50 °C). The actual temperature in the
furnace is changed without adjusting the displayed value.
Example:
-
In case of “underfiring”: raise the calibration value: +5 °C (instead of 0)
-
In case of “overfiring”: the value should be lowered: -5 °C (instead of 0)