Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
21
Vacuum pump
This parameter allows to choose between 3 modes of operation of the vacuum pump.
To modify, press the
▼
key to go to the drop-down menu and confirm your choice.
1.
OFF: the pump is switched off during the cycle
2.
ON: continuous operation within its capacity
3.
REG: active until the required vacuum level (8 mbar), then switches to regular mode via
intermittent operation
Example: vacuum level set at 60 mbar
The vacuum reaches 52 mbar (setting -8 mbar), the pump stops and restarts when the vacuum level
reaches 68 mbar (s8 mbar)
ATTENTION: in REG mode, if the vacuum level (- 8 mbar) cannot be reached, considering the performance
of the pump, the operation will continue without interruption within the running cycle
Vacuum start
Always ≥ preheating temperature and ≤ vacuum-off temperature
Vacuum stop
Always ≥ vacuum-on temperature and ≤ final temperature
Vacuum level (0 to 500 mbar)
This parameter is active in the REG mode of the vacuum pump, only
Vacuum hold (0’00’’ to 59’59’’)
It is the holding time for the vacuum at the final temperature. Programmable in minutes et seconds.
Value: 0 to 59 minutes and 59 seconds
Final temperature (maxi: 1050°C, peak: 1150°C)
Always ≥ preheating temperature and ≤ maximum temperature (1050°C, and peak temperature 1150 °C)
Hold time (0’00’’ to 59’59’’)
It is the holding time at the final temperature. Value: 0 to 59 minutes and 59 seconds
Internal cooling time (0 to 60 minutes)
After the holding time, it indicates the time during which the chamber remains closed during “natural”
cooling. Value: 0 to 60 minutes. If the value = 0 : next phase will start