Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
20
5.4.1.
PROGRAM PARAMETERS (ZONE 3)
IMPORTANT: if the numeric value of the parameter exceeds the minimum/maximum limits, a message
will indicate the limits to be observed. If these parameters are not modified, the indicated limits will be
taken into account when proceeding to the next parameter
Changes the field numerical:
-
Press the field. The value is displayed in green. Carry out the changes with the keypad numerical
-
Confirm and press key
: pass to following parameter
Program’s name
Select the corresponding field (white) to enter or change the name of the programme. This can be done
via the alphanumeric keypad that appears. To confirm or cancel, use the button in the lower left of the
screen
Drying rate (20°C to 150°C/min)
Indicates the rate of increase between the standby temperature and the preheating temperature.
Value: 20°C/min to 150°C/min
Preheating temperature (20°C to 1.000°C)
Always ≥ stand-by temperature and ≤ final temperature. Value: 20°C to 1 000°C
Close furnace time (0 to 60 minutes)
-
Set at 0: the furnace will close in one continuous movement when the preheating temperature is
reached
-
If the value is > 0: the tray will move in 6 equal stages during the selected time
Press
for close the furnace before the end of each stage. When press
, the tray stop in position
chosen by user. Press again
to movement will continue
Preheating time (0 to 60 minutes)
Within a running cycle and with the furnace closed, it indicates the holding time at the preheating
temperature
.
Value: 0 to 60 minutes
Rate (20 to 150°C/min)
Indicates the rate of increase between the preheating temperature and the final temperature
Value: 20°C/min to 150°C/min