6.2.
PROTECTION
OVERVOLTAGE
This furnace is protected against over-voltage by a system of active and passive filters. As a precautionary
measure, we recommend to save all programmes on the USB stick supplied with the unit.
OUTLET VOLTAGE VARIATION
Within a limit of ±20 volts, the furnace self-corrects the variation and maintains the same accuracy.
INTERRUPTION OF POWER SUPPLY
After a power failure of less than 15 seconds, the running cycle continues. When the furnace is switched
on again, it is rebooted and the graphics display reappears. No noticeable effect as the programme
proceeds.
If the power failure lasts more than 15 seconds, the cycle is interrupted as the duration of the incident will
substantially change the performance of the programme. The chamber opens and a message indicates that
a power interruption has occurred: MAINS FAILURE (see chapter 6.3)
OVERHEATING PROTECTION
If the temperature accidentally exceeds the programmed temperature by more than:
·
50°C, for a parameter setting > 800°C, or
·
70°C, for a parameter setting < 800 °C,
the running cycle is interrupted. The chamber opens and the message HEATING FAILURE is displayed (see
chapter 6.3)
DATA PROTECTION
The control slot contains a SD mini card memory. It ensures that all furnace data are saved and can be
inserted into another slot in case of irreversible damage of the CPU-card.