42
In the 5th stage, the gain adjustment is made
automatically. This step may take some time to
complete. When the stage is completed, the
“
NEXT
”
button becomes active. Clicking the button will
move to the next step.
In the 6th step, offset adjustment is made
automatically. When the stage is completed, the
“
NEXT
”
button becomes active. Clicking the button
will move to the next step.
In the 7th stage, the open circuit impedance
adjustment is made automatically. When the stage is
completed, the
“
NEXT
”
button becomes active.
Clicking the button will move to the next step.
In the 8th step, probe calibration is performed. The probes of the device are connected to the probe-1
input and the terminals are short-circuited. When short-circuited, the impedance value is seen and the
“
NEXT
” button b
ecomes active. Clicking the button will move to the next step.