https://ucore-electronics.com
10
Vertical offset can be adjusted in three ways.
•
When the channel is active, the signal is moved up and down the graphics area by holding and
dragging the touchscreen pen (left button if using the Mouse).
•
When the Rotary-1 button is pressed once, the offset adjustment mode is entered. After switching
to offset mode, the offset level is changed by rotating Rotary-1.
•
It is increased and decreased with the “
Inc. Offset
” and “
Dec. Offset
” buttons in the submen
u of
the
CH-1
menu. In addition, when the “
Rst. Offset
” button in the submenu is
clicked, the offset
value is set to 0V.
Coupling adjustment:
The undesired signals can be filtered out by setting the coupling mode. For example, the signal under test is
a square waveform with DC offset.
•
When the coupling mode is
“
DC
”
: the DC and AC components of the signal under test can both pass
the channel.
•
When the coupling mode is
“
AC
”
: the DC components of the signal under test are blocked.
Click the
“
CH-1
”
menu butto
n. The desired coupling can be selected with the “
DC-Coupl.
” and “
AC-Coupl
.
”
buttons in the submenu.
DC coupling
AC coupling