Page 76 of 85 Pages
5.
Press the STOP key to interrupt the operation, and exhaust
steam from the chamber.
6.
Wait until the pressure decreases to zero, only then can the
door be opened.
11.2.2
ASME-approved type safety valve
1.
Operate the sterilization cycle according to the manual.
2.
Allow a pressure of approximately 200 kPa (29-psi) to build
up in the chamber.
4.
Operate the safety valve:
5.
Pull the valve’s handle and verify that steam escapes from
the valve.
6.
Press the STOP key to pause operation, and exhaust steam
from chamber.
7.
Wait until pressure goes down to zero, only then can the
door be opened.