10.3
Safety
Protective equipment and clothes and other safety instructions should be
implemented in accordance with local and national regulations and/or
rules!
For proper sterilization, do not overload the chamber. Only autoclavable
products shall be used; please refer to the manufacturer instructions for
sterilization of unknown materials or instruments.
10.4
Loading
Correct loading of the autoclave is essential to successful sterilizing for
several reasons. Efficient air removal from the chamber and the load will
permit total steam penetration and saturation, and allow proper drainage
of condensate. Additionally, correct loading will reduce damage to packs
and their contents and maximize efficient use of the sterilizer.
For detailed loading instructions, see sec. 9 (Preparation before
sterilization)
10.5
Selecting the Program
1. Select the program.
Pressing the UP key selects the next program.
Pressing the DOWN key selects the previous program.
2. Make sure Verify that you chose the required program.
3. If the autoclave is equipped with a printer verify that a paper roll is
inserted in the printer. If not - insert as per para. 5.2.
Attention:
Selecting a program is possible only when the door is open.
10.6
Closing the door
1. Close the door.
2. Move the door handle (1) from position "unlocked" (3) to position
"locked" (2), see the figure above.
3. The open door symbol
is replaced by the message "System
Ready".
10.7
Starting cycle
1. Start the cycle by pressing the START/STOP key.
If your autoclave supports
CFR 21 part 11
standard,
perform the following procedure:
SELECT USER screen will be displayed: