Page 40 of 85 Pages
Program 14: Isothermal
For liquids when the manufacturer recommends autoclaving at 60
C
(140
F) with no drying.
Recommending for melting of agar,
pausterization etc.
Caution: This is not a sterilization program!
Nominal parameters default settings
Heating temperature: 60
C (140
F).
Heating time: 20 minutes.
Operations sequence:
Heating stage; steam enters the chamber and warms up until the
sterilization temperature 60
C (140
F) is reached inside the chamber.
Sterilization temperature is maintained constant for the preset
sterilization time.
Slow exhaust, steam is slowly exhausted from the chamber, until the
pressure equals the atmospheric pressure.