Page 70 of 85 Pages
1.
Enter the Enter Code screen by moving the cursor to your
username (Use
Up
and
Down
keys) and pressing
START/STOP key. The following screen will be displayed:
2.
0000 is displayed on the screen with the cursor blinking
on the right digit.
3.
To increase or decrease the digits, press the UP or
DOWN keys.
4.
Set your password, then move the cursor to Set
by pressing the START/STOP key.
When Set is blinking, press the UP or DOWN keys to return
to the program screen.
The autoclave starts performing sequence of operations. The actual
measured values of pressure and temperature are displayed
continuously and printed every minute at
STE
stage, and every 5
minutes at the other stages. The phase in progress is displayed at the
right side of the upper line as
WATER
,
HEAT
,
STER
.,
EXH
., and
DRY
.
If the operator presses the
START
key and the door is not completely
closed, the process will not start and the
DOOR
indicator will flash
twice then turn off and the buzzer will sound four times.
Caution!
Do not touch the strainer’s cover, mounted on the exhaust line,
during and short after operation.
Touching the hot strainer’s cover may cause severe injuries.