Page 41 of 85 Pages
6
CHECKING AND CHANGING PARAMETERS AND OTHER DATA
The control system prevents changing programs if the door is closed. This
protection is intended to avoid program changes if the autoclave is loaded. If the
operator for example inserts the load into the chamber, closes the door and leaves
the room and another operator/user tries to change the program, the operator/user
will not be able to do this unless the door is opened and the load inside the
chamber can be seen.
6.1
Directories and subdirectories
The operator (Admin) may perform the following:
DIRECTORY
SUBDIRECTORY
System Parameters
Print Rate All
Print Rate Sterilization
Screen Saver
History
View old cycle history
Export history to USB
Maintenance
Set date and time
Export gain offset to USB
Printer test
Print all gain and offset
Advanced options
Start cycle by clock
Export all settings to USB device
Version information
View current version information
View factory default version information
View previous version information
6.2
Entering the main menu
1.
Enter the SELECT USER screen by pressing the UP and DOWN
keys simultaneously.
To exit the SELECT USER screen move the cursor to Exit by
pressing UP or DOWN keys and then press START/STOP key
SELECT USER screen will be displayed.