Page 83 of 85 Pages
Message /
Symbol /
Problem
Failure Description
Corrective Action
Supply water
error (digitat
input option)
This message is displayed in case of
a city (tap) water supply
malfunction.
Check and fix the city (tap)
water supply.
Compressed
air supply
error (digitat
input option)
This message is displayed in case of
a compressed air supply
malfunction.
Check and fix the air supply.
No Water
This message is displayed if the
electrode in the chamber did not
sense water within the preset time.
1. check and fix the mineral
free water supply.
2.check and clean the water
inlet filter.
3. Clean the water level
electrode.
Analog Input
Error
This message is displayed when any
Temperature sensor or Pressure
sensor is disconnected or out of
range.
Call for service.
Chamber
temperature
not in range
This message is displayed if the
temperature in the chamber is too
high or too low from the normal
range.
Call for service.
Chamber
pressure not
in range
This message is displayed if the
pressure in the chamber is too high
or too low from the normal range.
Call for service.
I/O Card
Failed
This message is displayed if I/O
card is faulty (both while cycle is
running or not).
Call for service.
I/O card is
not connected
This message is displayed if I/O
card is disconnected (both while
cycle is running or not).
Call for service.