Page 11 of 85 Pages
3.4.2
Other utilities
Compressed Air (ELVC only)
1/2" 3-4 Bar (44-58 psi)
Tap water
1/2", 2-3 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp.
of 80°C (176°F)
Attention:
The electrical net must be protected with a current leakage safety relay.
The electrical network must comply with local rules or regulations.
3.5
Construction
The main parts of the autoclave are made of materials as indicated below:
Chamber and door are made of stainless steel
Trays are made of stainless steel.
Door handle is made of hard plastic material, which is safe to touch
and thermo-insulated.
3.6
Environment Emission Information
1.
The peak sound level generated by the autoclave is less than 70 dBa
with background noise of 60 dBa.
2.
The total heat per hour transmitted by the autoclave is < 100 W/h for
all models.
3.7
Directives and Standards
Every autoclave meets the provisions of the following Directives and is
constructed in compliance with the following Standards:
3.7.1
Technical Directives
1.
Pressure Equipment Directive 97/23/EC.
2.
Council Directive for low voltage equipment 73/23/EEC.
3.
Electromagnetic Compatibility Requirements Directive
89/336/EC.
3.7.2
Technical Standards
1.
AAMI/ANSI ST 8.
2.
ASME Code Sec. VIII for Pressure Vessels.
3.
Pressure Equipment Directive 97/23/EEC.
4.
EN 285:2006 ― Sterilization, Steam sterilizers , Large
sterilizers
5.
IEC/UL 61010A-
1 ― electrical equipment for laboratory
use; general requirements.
6.
IEC 61010-2-
040 ― particular re
quirements for sterilizers
and washer disinfectors used to treat medical materials.
7.
ISO 17665 ― 1:2006 Sterilization of health care products
moist heat.