Program 4: Liquid B
–
Waste
For liquids when the manufacturer recommends autoclaving at
temperatures of 121
C (250
F) for 30 minutes, such as liquid waste.
Note: both PT100 temperature sensors must be inside the bottles
For proper sterilization, fill the bottles with approximately the same
amount of liquid.
Nominal parameters default settings
Sterilization temperature: 121
C (250
F)
Sterilization time: 30 minutes
Operations Sequence
(PV, pre-vacuum model only) Air removal stage; four vacuum pulses
removes air by drawing the chamber down to 20kPa.
Heating stage; steam enters the chamber and warms up until the
sterilization temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Slow exhaust; the steam is slowly exhausted from the chamber, until it
reaches the required end temperature (approx. 85ºC) and the pressure
equals the atmospheric pressure.