Page 27 of 51
4.2
Displayed Error and Operational Messages
Message
Message Description
Corrective Action
Pt1 out
This message is displayed when
the Chamber Temperature sensor
Pt1 is disconnected or out of range.
Call for service.
Pt2 out
This message is displayed when
the Chamber Temperature sensor
Pt2 (additional safety feature) is
disconnected or out of range.
Call for service.
Low
Temperature
This message is displayed when the
sterilization temperature wasn’t
reached.
•
Check that the autoclave
is not overloaded (see sec.
2.6).
•
Check for steam leakage
from the door.
•
If the problem persist, call
the technician for service.
High
Temperature
This message is displayed
if temperature rises 3°C above
sterilization temperature during the
sterilization stage for 10 seconds.
Call for service.
No Water
This message is displayed if during
any cycles there is no water in the
chamber.
•
Check the mineral free
water supply is on.
•
If the problem persist, call
for service.
High
Pressure
This message is displayed if the
Chamber Pressure rises 20kPa
above sterilization pressure for 5
seconds during the sterilization
stage.
Call for service.
Low
Pressure
This message is displayed if
Chamber Pressure drops bellow the
sterilization pressure for 5 seconds
during the sterilization stage.
Call for service.
Manual Stop
This message is displayed if the
STOP key is pressed and the cycle
is aborted.
N/A
Low Heat
This message is displayed during
the heating stage, if the system
cannot reach the required chamber
temperature within 20 minutes.
•
Check that the autoclave
is not overloaded (see sec.
2.6).
•
Check for steam leakage
from the door.
•
If the problem persists, call
for service.
Press Sensor
Cutout
This message is displayed if
pressure sensor 1 of the chamber is
disconnected or if the pressure is
out of range during Air-Removal
and Heating stages.
Call the technician