Page 28 of 51
Message
Message Description
Corrective Action
No Res Src.
This message is displayed if there
is no water supply.
Check the connections to
the water supply.
Temp2 High
This massage is displayed, and
door opening is prevented, if
temperature sensor 2 senses a
higher temperature than the pre-set
end temperature.
Wait until the autoclave
cools to "end-temp".
Door is Open
Close the door to operate
the autoclave.
You can run a new cycle
only when the door is close
and the icon disappears.