Page 9 of 51
2.6
Loading Capacities
Maximum load
Model
liquid
instrument
2540
10 liters
22 lb
3850/70
20 liters
44 lb
2.6.1
Erlenmeyer Flasks
SIZE
2540
3850
3870
250 ml
2 x 5
3 x 12
4 x 12
500 ml
2 x 4
2 x 7
3 x 7
1000 ml
1
2 x 4
2 x 4
2000 ml
1
3
2 x 3
3000 ml
1
2
2 x 2
5000 ml
1
1
1
2.6.2
Medium Flasks (Schott)
SIZE
2540
3850
3870
250 ml.
2 x 8
3 x 18
4 x 18
500 ml
2 x 5
2 x 11
3 x 11
1000 ml
4
2 x 8
2 x 8
2000 ml
1
4
2 x 4
5000 ml
1
2
2
10000 ml
1 1
2.7
Utility
Utility
Value
Power
1Ph, 208V/60Hz
2540
Recommended
Circuit Breaker
20A
Power
3 Ph, 208V/60Hz
Electrical
data
3850
/7
0
Recommended
Circuit Breaker
3x20A
Compressed Air (ELVC only)
1/2", 60-120psi (4.1-8.3 Bar), 60 l/min
Tap water (ELVC only)
1/2", 30-45 psi (2.1-3.1 Bar)
Mineral free water
1/2", 30-45 psi (2.1-3.1 Bar)
Drain
Must be able to withstand a minimum
temp. of 190°F (88°C) min.
Attention:
The electrical net must be protected with a current leakage safety relay.
The electrical network must comply with local rules or regulations.
Содержание 2540 ELV
Страница 2: ......
Страница 15: ...Page 13 of 51 OVERALL DIMENSIONS DRAWING FOR 2540 ELV ELVC...
Страница 26: ...Page 24 of 51 4 KEYBOARD Keys and Display Screen Keypad Power On Off Switch...