Page 40 of 51
•
Select the required program by means of the keys
UP
,
DOWN
and
START/STOP .
UP
key: next program
DOWN
key: previous program
START/STOP
key: start the selected program
8.5
End of the Cycle
•
At the end of the cycle the message “CYCLE END!” is displayed on
the screen and the buzzer sounds one beep continuously during 3
seconds.
•
In case the cycle has failed, the message “CYCLE FAIL!” is displayed
on the screen, the buzzer sounds intermittently 3 beeps and an Error
Message
is displayed on the screen. (For corrective action see
Displayed Error and Operational Messages sec 4.2) Pressing the
START/STOP
key will erase the fail message.
8.6
Unloading
•
On completion of the cycle, open the door to remove the sterilized
material from the autoclave. Rotate the handle of the tightening bolt
counter-clockwise to the end then lower the bolt at a right angle,
parallel to the front cover. Lift the door to a vertical position.
Caution
Before opening the door, verify that there is no pressure in the
chamber (a chamber pressure gauge is located on the autoclave's side
cover).
Open the door slowly to allow steam to escape. DO NOT put your
head or any other part of your body above the chamber. Doing so
can cause
injury from hot steam.
•
Take the load out of the sterilizer immediately upon completion of the
cycle. Do not remove the load from the basket until its temperature has
returned to room temperature. Let the load cool down in an area without
air movement (air conditioning, etc.) and with minimum people passing
by to avoid the possibility of touching the hot load. Do not touch the hot
load since a hot load will absorb moisture and, therefore, may absorb
bacteria from your hand. Perform a visual inspection to ascertain that
sterilizing indicators have made the required color change, and that the
load is dry.
•
The load shall be rejected if:
a.
The package has been compressed.
b.
The package is torn.
c.
The load is suspected to be wet.
d.
The load fell on the floor.
e.
Condensed drops can be detected on the load.
To avoid injuries use heat resistant gloves while unloading the
autoclave.