Check the instructions of the item manufacturer as to the proper
procedure for sterilizing each item.
9.
Items must be sterilized in an open position. Surfaces that are hidden
because the item is in a closed position will not be exposed to the
steam and will not be sterilized.
10.
Use single-use wraps once only and discard after use.
11.
Materials used for wrapping and packing, including materials used for
inner wraps, shall be compatible with the item being packed and the
sterilizing method selected.
12.
Wrapped instrument should be placed in material which will allow
steam penetration and promote drying, such as autoclave bags,
autoclave paper, or muslin towels.
13.
If necessary, disassemble or loosen multi-part instrument prior to
packaging to permit the sterilizing agent to come into contact with all
parts of the instrument.
14.
Tilt items on edge that are prone to entrap air and moisture, e.g.
hollowware, so that only minimal resistance to air removal, passage of
steam and condensate will be met.
15.
Load the basket loosely to capacity
16.
Place a sterilization indicator strip in each basket.
17.
Once a week, use a biological spore test indicator in any load to make
sure sterilization is performed.
18.
Make sure that all instruments remain apart during the sterilization
cycle.
7.2
Textiles
1.
For all loads not containing liquids, insert the two temperature sensors
(1) into the sensor holder (2) attached to the autoclave’s door (3) (see
picture 7.1 #1).
2.
Launder textile wraps prior to sterilization, but DO NOT use bleach.
3.
Tilt textile packs on edge to provide minimal resistance to the removal
of entrapped air.
4.
Wrapped textiles should be placed in material which will allow steam
penetration and promote drying, such as autoclave bags, autoclave
paper, or muslin towels.
5.
Materials used for wrapping and packing, including materials used for
inner wraps, shall be compatible with the item being packed and the
sterilizing method selected.
6.
Load the basket loosely to capacity.
7.
Place a sterilization indicator strip in each basket.
7.3
Empty containers”
1.
For all loads not containing liquids, insert the two temperature sensors
(1) into the sensor holder (2) attached to the autoclave’s door (3) (see
picture 7.1 #1).
2.
Place empty containers upside down to prevent accumulation of water.