PREPARATION BEFORE STERILIZATION
The purpose of packaging and wrapping of items for sterilization is to provide a
protective
barrier against recontamination after items are removed from the
sterilizer. Packaging and wrapping materials should permit the removal of air
from the pack, penetration of the sterilizing water vapor into the pack and removal
of the sterilizing vapor.
The basic principle determining the size, mass and contents of instrument and
hollowware packs is that the contents are sterile and dry immediately on
completion
of
the
drying
cycle
and
removal
of
the pack from the sterilizer chamber.
Instruments to be sterilized must be free from all residual matter, such as blood, or
organic tissue. Instruments must also be dry and free from mineral deposits. Such
substances may cause damage to the to instruments themselves or the sterilizer.
If the unit is equipped with a printer, check if a new roll of paper is needed.
7.1
Instruments
1.
For all loads not containing liquids, insert the two temperature sensors
(1) into the sensors holder (2) attached to the autoclave's door (3) (see
picture below).
2.
Clean instruments immediately after use to remove any residue. It is
recommended that all instruments be ultrasonically cleaned using
Tuttnauer’s CLEAN AND SIMPLE enzymatic cleaning tablets or
other suitable solution.
3.
After cleaning, rinse instrument under tap water for 30 second and pat
or air dry to remove residual minerals. If your tap water has a high
mineral content rinse a second time in a bath of distilled water to
remove minerals.
4.
Follow the instrument manufacture’s instructions on the use of
products for cleaning and lubricating instruments that have been
ultrasonically cleaned.
5.
Be sure that instruments of dissimilar metals (stainless steel, carbon
steel, etc.) are separated. Carbon steel instruments should be bagged
or placed on autoclavable towels and not be allowed to come in
contact with the stainless steel baskets (mixing will result in the
oxidation of these metals).
6.
Do not place materials to be sterilized against the chamber’s wall.
Place the material only in the basket.
7.
When using a paper / plastic bag, the plastic side should always be
down.
3