Load Capacity
68
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
18.1.5
Operating table top with Carbon 1200 table top segment
[34] Carbon 1200 table top segment
1. No section segment may be attached to the head end of the
operating table top.
2. A Carbon 1200 table top segment [34] is attached at the foot
end of the operating table top.
3. Patient orientation on the operating table top is as normal or
inverse. The permissible patient weight depends on the patient
position.
Permissible load:
Attention, additional accessories reduce the permissible patient
weight.
18.1.6
Operating table top with extension unit
[35] Extension unit
1. The extension unit [35] is attached at the foot end of the
operating table top.
2. Patient orientation on the operating table top is as normal.
Permissible load:
18.1.7
Operating table top with shoulder chair
[36] Shoulder chair
[37] Head positioning accessory
1. No section segment is necessary at the head end of the
operating table top.
2. The shoulder chair [36] with the head-positioning
accessory [37] is fastened to the foot end of the operating
table top.
3. Patient orientation on the operating table top is as inverse.
353 lbs
(160 kg)
Patient weights of up to 353 lbs (160 kg) in the normal
patient position are possible with restriction of the
operating table functions.
298 lbs
(135 kg)
Patient weights of up to 298 lbs (135 kg) in the inverse
patient position are possible with restriction of the
operating table functions.
353 lbs
(160 kg)
A patient weight of up to 353 lbs (160 kg) is possible
without limiting the operating table functions.
Содержание 1604786
Страница 1: ...TruSystem 7000 Operating Table Instruction manual...
Страница 4: ...4 Instruction manual Operating table TruSystem 7000 1 730 263 11 2016...
Страница 109: ......