Attaching the Section Segment to the Operating Table Top
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
55
14
Attaching the Section Segment to the Operating Table Top
14.1
Section segment without additional locking
A typical section segment without additional locking is the light one-
part leg section (closed section segment).
Attaching the section
segment
1. For a closed section segment, bring the joints on the operating
table top to the same angular position.
2. Insert the section segment with the hooks [A] obliquely from
above, from the front or obliquely from below into the fixing
points of the operating table top.
3. Pull the section segment [A] away from of the operating table
top. The hooks must be guided precisely into the recesses of
the fixtures.
4. Check that the section segment is securely attached to the
operating table top. The section segment can fall from the
operating table top if the hooks are not locked securely in
place. The section segment is not secured to the operating
table top by other means.
CAUTION!
Risk of personal injury. Risk of injury due to improper
attachment!
Unlocked section segments can fall away from the operating table
top and cause injury to staff or patients.
– Follow the user manual of the section segment.
– Check that the section segment is securely attached to the
operating table top before use.
– Never attach or detach several section segments
simultaneously.