Power Supply
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
45
Connecting the
operating table to
the power supply
Flip up the connector socket cover [15] in the running
gear [13] and insert the power cable plug [19] into the socket
as shown [15]. The connection location is identified by the
symbol for the line power connection.
3. Route the cable such that no one can trip or fall over it.
Plug the connector of the power cable into a grounded power
socket in the room. The grounded power socket must be
outside of an area where there is a risk of explosion.
There is a risk of explosion in conjunction with flammable
mixtures of anesthetics and air.
4. Indicators [i1] and [a3] on the column keypad light up.
After a few seconds, an audible signal sounds. The operating
table is only ready for use once the signal tone has sounded.
If the operating table was switched off, it will switch on
automatically when plugged into the wall outlet.
1.
DANGER!
Risk of death from damaged power cables!
There is a risk of death by electric shock if the
power cable is damaged. Check the power
cable and remove from use if it has been crushed
or if the insulation is damaged.
2.
WARNING!
Risk of explosion due to flammable gas
mixtures!
There is risk of explosion in conjunction with
flammable mixtures of anesthetics and air if the
power cable is first plugged into the socket and
then into the operating table. The sequence for
connecting the power cable must be strictly
followed.