Safety
14
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
WARNING!
Risk of personal injury.
When used on an electrically conductive floor, the
operating table may be used only if connected to an
equipotential bonding cable. The equipotential
bonding cable connection location on the operating
table is in accordance with IEC 60601–1.
CAUTION!
Risk of personal injury and material damage.
Ensure that the power cable and equipotential bonding cable are
not crushed or driven over. Stop using cables if damaged. If there
are concerns regarding the safety of the power or equipotential
bonding cable, use the internal power supply (battery) only until the
corresponding cable can be replaced.
Remove cables prior to relocating the table.
CAUTION!
Risk of personal injury.
The anti-static properties of the operating table top comply with the
legal requirements. Meeting this requirement requires the use of the
pads supplied with the product, which must be properly secured.
NOTICE
Risk of material damage to the electronics due to
condensation.
There is a risk of condensation forming during transport, which
could potentially damage electronic components. After delivery of
the operating table, it is important to wait a certain period of time
before it is switched on or connected to an external power source
(at least 12 hours is recommended). Place the operating table in
an environment with the same ambient temperature and relative
humidity as the room where the table will be used (see page 96).
The batteries must be completely recharged after a significant
change in temperature or humidity.