Malfunction Caused by other Devices
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
43
7
Malfunction Caused by other Devices
Portable and mobile RF communications equipment may interfere
with the medical device.
Examples of potential sources of interference:
– Electronic control gear (ECG) for fluorescent lamps
– RF surgical devices
– wireless remote controls for other devices (for example, monitor)
– very bright interior lights
WARNING!
Risk of personal injury. Hazard to patients caused by
malfunctions of the operating table!
It is possible that other medical or non-medical devices in the room
may use the same frequency range as the operating table or the
wireless remote control. Thus, the devices can cause the
malfunction of the operating table. Note the operating frequencies
of various pieces of equipment.