Reset
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
49
If the user presses the
C
ANCEL
key [i30] or the
▸
Cancel
◂
button, the
operating table shuts down in half an hour (after the 4 hours
elapse).
If the user does not react, the audible signal goes off a total of
12 times. The message remains on the remote control display. The
operating table shutdown time is not extended. The operating table
also shuts down after the 4 hours elapse.
It may be necessary for an operating table in battery mode to be
switched on again in the interim for longer surgeries without the
operating table controls being used.
1. Disconnect the operating table from the power supply (see
section 8.2 on page 44).
2. Press key [i2] on the column keypad or remote control for
more than 2 seconds. An acoustic signal sounds shortly before
shutdown.
11
Reset
If the operating table reacts unexpectedly to a control, switch off the
table and switch it back on. If it is not possible to switch off the table,
press the Reset button. The reset button [23] is located at the head
end on the column base, directly below the cladding. Remove the
cover cap [24] (made of plastic) and press the button [23] briefly
using a pointed object (e.g., a ballpoint pen). Then switch the
operating table on again. With an external power supply, the
operating table switches on again automatically.